The Research of Skin Imaging Technology with High Frequency Ultrasound
10.3969/j.issn.1671-7104.2013.06.004
- VernacularTitle:高频超声皮肤成像技术的研究
- Author:
Jun YANG
1
;
Jianjun JI
;
Yuejie LI
;
Yanqun WANG
;
Yawei LIU
;
Yunhong JI
;
Xuedong SONG
Author Information
1. 中国医学科学院北京协和医学院生物医学工程研究所
- Keywords:
high-frequency ultrasound;
skin ultrasound imaging;
linear scan
- From:
Chinese Journal of Medical Instrumentation
2013;(6):398-400
- CountryChina
- Language:Chinese
-
Abstract:
Objective Developing a high-frequency ultrasonic skin imaging system to obtain the high resolution ultrasonic image of the skin. And further analyzing the ultrasonic images of skin to explore the imaging characteristics of skin structure and then explore the value of high-frequency imaging in the application of skin diagnosis. Methods 50 MHz single element ultrasonic transducer, mechanic linear scanning method is used in the imaging system. The resolution and the ability of recognize the skin issue is verified by linear target scanning and clinical trials. Results Both the axial and lateral resolution of the system reaches 50 mm. The subtle structure of normal skin tissue is clearly visible. Some diseases have obvious appearance in the image. Conclusions 50 MHz ultrasonic skin imaging system is of high resolution and is valuable to skin structure detect and disease diagnosis.