CT Scan Optimization in SPECT/CT Equipment of Nuclear Medicine
10.3969/j.issn.1671-7104.2013.04.021
- VernacularTitle:核医学SPECT/CT设备的CT扫描方案优化
- Author:
Yuankai WANG
1
;
Zhuang LIU
Author Information
1. 复旦大学附属华山医院核医学科
- Keywords:
tomography;
X-ray computer;
radiation dose;
exposure time;
program optimization
- From:
Chinese Journal of Medical Instrumentation
2013;(4):308-310
- CountryChina
- Language:Chinese
-
Abstract:
Objective In order to reduce patients’ radiation doses and decrease the exposure time of the CT Tube. By summarizing experiential methods from the previous practice work, we hope to draw an optimized scheme of CT scanner. Methods We analyzed the continuous clinical examples in four months, changed the different levels of parameters slowly, optimized acquisition conditions and retrospected the CT radiative dose. In the meantime, we counted the monthly CT Bulb exposure time, exposure time per capita tube data and so on. Results Compared with former result, the average CT radiative dose per person is decreased 84.72%, which conventional bone, the smal joints of the bone, the brain and the heart radiation reduced respectively 83.82%, 88.80%, 86.78% and 84.49%, and the differences are statistical y significant. The exposure time of optimized CT Tube is 2 031 s. Conclusions Using the CT scan optimization scheming has great efforts in the practice of clinical nuclear medicine, can reduce the exposure time of the CT Tube and improve the lifetime of the bulb. Therefore, it can be widely used.