Establishment and Evaluation of Intraocular Lens Loop Support Force Test Platform
10.3969/j.issn.1671-7104.2013.04.016
- VernacularTitle:人工晶状体襻支撑力测试平台的建立及评价
- Author:
Li LIU
1
;
Linnan KE
;
Shuo WANG
;
Ying WANG
;
Haiping REN
Author Information
1. 中国食品药品检定研究院
- Keywords:
intraocular lens;
loop;
support force
- From:
Chinese Journal of Medical Instrumentation
2013;(4):291-293
- CountryChina
- Language:Chinese
-
Abstract:
Objective Establish the test platform of the intraocular lens loop, and the platform was evaluated through the experiment. Methods The intraocular lens loop test platform is made up with three models. The different intraocular lens haptics support force can be completed by replacing different sample holder model. Results The standard deviation and the coefficient of variation were calculated through the result of the fifteen samples. The standard deviation was 0.04 mN, and the coefficient of variation was 0.66%. The two values were in the acceptable range. Conclusions The platform was so stabilizing that it could be used to test support force of IOL loop. The different shapes of IOL could be tested on the platform through the replacement of the holder model.