Ideal current of injury and R-wave sensingvalues for identifying optimized placement of right ventricular active-fixation pacing leads
10.3760/cma.j.issn.0253-3758.2016.05.008
- VernacularTitle:联合损伤电流与R波感知优化右心室主动固定起搏导线的植入
- Author:
Jianhua CHEN
1
;
Feilong ZHANG
;
Lin FAN
;
Xuehai CHEN
;
Weiwei WANG
;
Fayuan FU
;
Lianglong CHEN
Author Information
1. 福建医科大学附属协和医院心内科福建省冠心病研究所
- Keywords:
Pacemaker,artificial;
Electrocardiography;
Current of injury
- From:
Chinese Journal of Cardiology
2016;44(5):406-410
- CountryChina
- Language:Chinese
-
Abstract:
Objective To define the ideal COI and R-wave sensing values for optimized placement of right ventricular active-fixation pacing leads.Methods Three hundred and six patients underwent right ventricular active-fixation lead implantation in our department from January 2013 to December 2014 were enrolled in this prospective pilot cohort study.According to the times needed for successful implantation,patients were divided into group SP (success poisoning after single procedure,n =229) and group MP (success poisoning after multiple procedures,n =77).Current of injury (COI) was characterized as the magnitude of ST-segment elevation.Pacing parameters were measured up to 10 minutes after lead fixation.Results (1) COI at 1 minute (COI1,(9.34 ± 2.44) mV vs.(3.19 ± 1.36) mV,P < 0.001) and 10 minutes (COI10,(4.99 ± 2.04) mV vs.(1.91 ± 0.62) mV,P < 0.001) after lead fixation was significantly higher in SP group compared to MP group.R-wave sensing measured at 10 minutes after lead fixation (R10 min) was also significantly higher in SP group than in MP group ((14.2 ± 4.5) mV vs.(5.4 ± 0.7) mV,P < 0.001).Pacing threshold measured at 10 minutes after lead fixation was significantly lower ((0.65 ±0.18)V vs.(1.90 ±0.40)V,P<0.001) in SP group than in MP group.(2) The optimized placement of pacing lead was related with COI1 (OR =0.247,95% CI0.098-0.623,P =0.003) and R10min (OR =0.081,95 % CI 0.015-0.428,P =0.003).(3) The cut-off value of COI1 was 4.77 mV (sensibility 0.97,specificity 0.95),the cut-off value of R10 min was 7.25 mV (sensibility 0.96,specificity 1.00) for optimized lead fixation.After 3 months follow up,pacing threshold was (0.68 ± 0.19)V,R-wave sensing was (12.1 ± 3.6)mV,and pacing impedance was (543 ± 107) Ω for right ventrieular active-fixation pacing leads.Conclusions COI1 > 4.77 mV and R10 min > 7.25 mV are the ideal parameters for identifying optimized placement of right ventricular active-fixation pacing leads.