Study on the matching feature of current parameters for inducing contraction of orbicularis oris muscle by artificial facial nerve in rabbit with peripheral facial paralysis
10.3760/cma.j.cn115330-20190917-00582
- VernacularTitle:人工面神经诱发面神经麻痹兔口轮匝肌收缩电流参数匹配的研究
- Author:
Dongyue XU
1
;
Yajing SUN
;
Ningjun ZHAO
;
Dan LUO
;
Yanxia ZHAO
;
Keyong LI
Author Information
1. 内蒙古赤峰学院附属医院耳鼻咽喉科 024050
- Keywords:
Facial paralysis;
Artificial facial nerve;
Orbicularis oris muscle;
Electric stimulation
- From:
Chinese Journal of Otorhinolaryngology Head and Neck Surgery
2020;55(6):615-619
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To explore the optimal matching of the current parameters about artificial facial nerve (AFN) for inducing contraction of the orbicularis oris muscle (OOM) in rabbit with peripheral facial paralysis.Methods:According to the combinations of different stimulus parameters, at different time points, we used AFN to induce contraction of the affected side OOM of the paralyzed rabbit in waking state. We recorded the current amplitudes of the threshold stimulation and peak stimulation under each combination, then compared the total charge of the stimulus consumption under different combinations.Results:Compared the total stimulus charge consumption of the AFN threshold stimulation and that in the peak stimulation respectively under different stimulus frequency and pulse width matching combinations, we found that the frequency, the pulse width and the interaction of the frequency and pulse width among different groups could affect the total charge, the differences were statistically significant ( P<0.05). Conclusions:When AFN stimulus frequency is 60 Hz and the stimulus pulse width is 100 μs, the output current intensity is relatively lowest and the total consumed charge is the lowest under the premise of effectively inducing the paralytic side OOM contraction. Thus we recommend this stimulus parametric combination as the optimal combination for meeting the low power consumption of AFN.