Determination of ertapenem in human plasma by UPLC-MS/MS
10.13699/j.cnki.1001-6821.2024.05.020
- VernacularTitle:用UPLC-MS/MS法测定人血浆中厄他培南的浓度
- Author:
Yi-Ran TAO
1
;
Bo-Yu LIU
;
Lei HU
;
Si-Lu LIU
;
Hai-Yan ZHAO
;
Xue-Cai XUE
;
Lin HUANG
Author Information
1. 北京大学人民医院药学部,北京 100044
- Keywords:
ertapenem;
liquid chromatography-tandem mass spectrometry;
therapeutic drug monitoring
- From:
The Chinese Journal of Clinical Pharmacology
2024;40(5):728-731
- CountryChina
- Language:Chinese
-
Abstract:
Objective A simple,sensitive and rapid ultra high performance liquid chromatography tandem mass spectrometry(UPLC-MS/MS)method was developed and validated for the determination of ertapenem in human plasma.Methods Using ertapenem-D4 as internal standard,the protein in plasma was precipitated with acetonitrile;chromatographic column:ACQUITY HSS T3(2.1 mm × 50.0 mm,1.8 μm);the mobile phase was 0.1%formic acid aqueous solution(containing 2 mmol·L-1 ammonium formate)-acetonitrile(0.1%formic acid),using a gradient elution;flow rate:0.4 mL·min-1,injection volume:1 μL,column temperature:45 ℃,the analysis time was 4.5 min,the scanning mode is positive ion selective reaction monitoring mode(SRM)with an electric spray ion source(ESI).The specificity,standard curve and lower limit of quantification,precision and recovery,matrix effect,dilution effect and stability were investigated.Results Ertapenem had a good linearity within 0.5-80.0μg·mL-1,and the standard curve was y=4.25 × 10-1x-2.64× 10-2(r2=0.999 0),the lower limit of quantification was 0.5 μg·mL-1,the relative standard deviation within and between batches is 1.39%-4.15%.The extraction recovery rate was 58.36%-64.57%,and the relative standard deviation of dilution effect was 3.30%,and the matrix effect was 99.71%-103.23%.The relative standard deviation of room temperature,repeated freeze-thaw,4 ℃,and long-term stability are all less than 10%.Conclusion The method is sensitive,rapid and specific,which is suitable for clinical monitoring of Ertapenem.