Abrasion of abutment screw coated with TiN.
10.4047/jap.2009.1.2.102
- Author:
Seok Won JUNG
;
Mee Kyoung SON
;
Chae Heon CHUNG
;
Hee Jung KIM
- Publication Type:Original Article
- Keywords:
screw loosening;
TiN-coating;
abutment screw
- MeSH:
Collodion;
Friction;
Joints;
Tin;
Titanium;
Weight Loss
- From:The Journal of Advanced Prosthodontics
2009;1(2):102-106
- CountryRepublic of Korea
- Language:English
-
Abstract:
STATEMENT OF PROBLEM: Screw loosening has been a common complication and still reported frequently. PURPOSE: The purpose of this study was to evaluate abrasion of the implant fixture and TiN coated abutment screw after repeated delivery and removal with universal measuring microscope. MATERIAL AND METHODS: Implant systems used for this study were Osstem and 3i. Seven pairs of implant fixtures, abutments and abutment screws for each system were selected and all the fixtures were perpendicularly mounted in liquid unsaturated polyesther with dental surveyor. After 20 times of repeated closing and opening test, the evaluation for the change of inner surface of implant and TiN-coated abutment screw, and weight loss were measured. Mann-Whitney test with SPSS statistical software for Window was applied to analyze the measurement of weight loss. RESULTS: TiN-coated abutment screws of Osstem and 3i showed lesser loss of weight than non-coated those of Osstem and 3i (P < .05, Mann-Whitney test). CONCLUSION: Conclusively, TiN coating of abutment screw showed better resistance to abrasion than titanium abutment screw. It was concluded that TiN coating of abutment screw would reduce the loss of preload with good abrasion resistance and low coefficient of friction, and help to maintain screw joint stability.