Analysis on the tracking evaluation of WS 582-2017 Specifications for Testing of Quality Control in X and γ Ray Stereotactic Radiotherapy System in Jiangsu Province
10.20001/j.issn.2095-2619.20240618
- VernacularTitle:WS 582—2017《X、γ射线立体定向放射治疗系统质量控制检测规范》在江苏省跟踪评价结果分析
- Author:
Xingjiang CAO
1
;
Wei CHEN
;
Xiangyong FAN
;
Jin WANG
Author Information
1. Jiangsu Provincial Center for Disease Control and Prevention, Nanjing, Jiangsu 210000, China
- Publication Type:Journal Article
- Keywords:
X and γ ray stereotactic radiotherapy system;
Quality control;
Standard tracking;
Questionnaire survey
- From:
China Occupational Medicine
2024;51(3):345-349
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the implementation status and problem of WS 582-2017 Specifications for Testing of Quality Control in X and γ Ray Stereotactic Radiotherapy System (hereinafter refer to as "WS 582-2017") in Jiangsu Province since it was published, and evaluate the scientific, rational, applicable, and operational aspects of the standard. Methods A total of 122 professionals from 54 medical institutions, radiation health technical service institutions, and health supervision institutions in Jiangsu Province who were involved in the use of WS 582-2017 were selected as the research subjects using a judgmental sampling method. A survey was conducted on the implementation of the standard, technical content, implementation effects, opinions and suggestions on the standard and implementation work. Results The WS 582-2017 awareness rate was 76.2%, and the training participation rate was 62.4%. Among the 93 individuals who were aware of WS 582-2017, 92.5% believed that the standard was scientifically rigorous, 73.1% believed that the standard was highly operational, 96.8% believed that the standard was consistent with relevant regulations and standards, 87.1% believed that the standard played an important role in promoting the optimization of radiation protection in radiotherapy, 89.2% believed that the standard played an important role in standardizing the quality control inspection of radiotherapy equipment, and 86.0% believed that the standard played an important role in improving the professional competence of relevant technical personnel. The tracking evaluation score of the standard in Jiangsu Province was 81.9 points. Conclusion The overall implementation of WS 582-2017 in Jiangsu Province is good, with good scientific and operational technical content and great implementation effects. But the content of the standard needs to be further improved, and training needs to be strengthened to increase the awareness of the standard.