Simultaneous determination of seven heavy metal elements in human blood and urine by inductively coupled plasma tandem mass spectrometry
10.20001/j.issn.2095-2619.20240814
- VernacularTitle:电感耦合等离子体-串联质谱法同时测定人血液和尿液中7种重金属元素
- Author:
Xiaoting LUO
1
;
Congmin LIANG
;
Ming DONG
;
Banghua WU
;
Junli SHAO
Author Information
1. Guangdong Province Hospital for Occupational Disease Prevention and Treatment, Guangzhou, Guangdong 510300, China
- Publication Type:Journal Article
- Keywords:
Inductively coupled plasma tandem mass spectrometry;
Mass spectrum interference;
Blood;
Urine;
Heavy metal element
- From:
China Occupational Medicine
2024;51(4):428-433
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish a method to simultaneously determinate cobalt, nickel, arsenic, molybdenum, silver, cadmium and lead in blood and urine using inductively coupled plasma tandem mass spectrometry (ICP-MS/MS). Methods The blood samples were diluted 20 times with a mixed solution of nitric acid (0.10% V/V) and Triton X-100 (0.02% V/V). The urine samples were diluted 10 times with nitric acid (1.00% V/V). Yttrium-89, rhodium-103, and lutetium-175 were used as internal standards to decrease matrix interference, and either On-Mass mode or Mass-Shift mode was used to decrease mass spectrometry interference, with detection by ICP-MS/MS. Results The linear ranges of the seven elements were 0.100-10.000 μg/L, with the correlation coefficient >0.999 9. The detection limits for the seven elements in the blood and urine were 0.003-0.021 and 0.003-0.031 μg/L, respectively, and the minimum quantification limits were 0.009-0.064 and 0.009-0.094 μg/L, respectively. The recovery rates were 96.64%-102.90% and 96.34%-104.50% for the blood and urine, respectively. The within-run relative standard deviations (RSDs) were 0.32%-2.33% and 0.25%-2.31%, and the between-run RSDs were 1.07%-3.81% and 1.30%-3.62% for blood and urine, respectively. The samples were stable at 4 ℃ for at least 14 days. Conclusion ICP-MS/MS is a simple, sensitive, and accurate method for rapid detection of seven heavy metal elements in the blood and urine of occupational hazard-exposed workers.