Comparison of the image quality between total-body PET/CT with 1 min acquisition and conventional PET/CT: a self-control study
10.3760/cma.j.cn321828-20221027-00322
- VernacularTitle:全身PET/CT 1 min采集图像与常规PET/CT图像质量比较:自身对照研究
- Author:
Runjun YANG
1
;
Xueqi CHEN
;
Haojun YU
;
Pengcheng HU
;
Hongcheng SHI
Author Information
1. 复旦大学附属中山医院核医学科、复旦大学核医学研究所、上海市影像医学研究所,上海 200032
- Keywords:
Image processing, computer-assisted;
Time factors;
Positron-emission tomography;
Tomography, X-ray computed;
Fluorodeoxyglucose F18
- From:
Chinese Journal of Nuclear Medicine and Molecular Imaging
2024;44(1):12-15
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To investigate whether the image quality of total-body PET/CT (TB PET/CT) with 1 min acquisition can meet the clinical diagnostic requirements.Methods:From May 2019 to September 2021, a total of 90 malignant tumor patients (60 males, 30 females, age 31-86 years) with primary lesions confirmed by pathological diagnosis in Zhongshan Hospital, Fudan University were respectively analyzed. All patients underwent conventional PET/CT (C PET/CT) scan with conventional clinical acquisition and TB PET/CT scan with 1 min acquisition after injecting 18F-FDG in random order. Paired t test or Wilcoxon signed rank test was used to analyze the image quality of these two scans. Results:SUV max of primary lesions in TB PET/CT group was significantly higher than that in C PET/CT group (15.9(7.9, 24.6) vs 12.5(5.8, 16.6); z=8.14, P<0.001), so were signal-to-noise ratio (SNR) of the blood pool, liver, muscles (9.3±3.0, 11.4(9.5, 14.2), 8.3(7.3, 10.1) vs 6.2±1.7, 9.4(7.7, 11.8), 6.0(4.9, 7.1)), tumor-to-blood pool ratio (TBR) (9.3(4.3, 14.8) vs 8.5(4.3, 11.1)), tumor-to-liver ratio (TLR) (6.7(3.0, 10.4) vs 6.1(2.9, 7.7)), tumor-to-muscle ratio (TMR) (23.2(11.5, 38.0) vs 18.3(9.6, 26.6); t=9.36, z values: 4.44-7.40, all P<0.001). Conclusion:The image quality of TB PET/CT scan with 1 min acquisition can meet the diagnostic requirements, and is better than the C PET/CT image quality with conventional clinical acquisition.