Clinical value of prospective ECG-gated high-pitch CT scanning in the diagnosis of pediatric congenital heart disease
10.3760/cma.j.cn112271-20230619-00200
- VernacularTitle:前瞻性心电门控大螺距CT检查对小儿先天性心脏病诊断的价值
- Author:
Minghua SUN
1
;
Yukun PAN
;
Ru WEN
;
Jiliang ZHANG
;
Ruigang XIE
;
Yinghui GE
Author Information
1. 阜外华中心血管病医院放射科 河南省心脏病影像医学重点实验室,郑州 451464
- Keywords:
Heart defects, congenital;
Tomography, X-ray computed;
Dual-source CT;
Radiation dose
- From:
Chinese Journal of Radiological Medicine and Protection
2024;44(3):228-232
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To investigate the clinical value of prospective ECG-gated high-pitch protocol scanning of third generation DSCT in the diagnosis of pediatric congenital heart disease (CHD).Methods:A total of 243 children with confirmed CHD who were expected to undergo surgical treatment were prospectively collected and randomly divided evenly into 3 groups, with first group for prospective ECG-gated high-pitch scanning in third generation DSCT (Flash 3rd), second group for prospective ECG-gated high-pitch scanning in second generation DSCT (Flash 2nd) and third group for prospective sequential scanning in third generation DSCT (Sequence 3rd). The SD value and SNR of aortic root and pulmonary artery of each child were recorded. The 5-point system is adopted with subjective scoring. Based on the result of operation, the diagnosis accuracy in 3 groups was analyzed. Results:The E values in Flash 3rd, Flash 2nd and Sequence 3rd group were 0.24 (0.19, 0.27), 0.11 (0.10, 0.14) and 0.44 (0.39, 0.48) mSv ( H=207.04, P<0.05), respectively. Subjective scores of group Flash 3rd and Sequence 3rd were significantly higher than that of group Flash 2nd [4 (4, 4) vs. 4(3, 4) vs. 3(3, 3), H=124.05, P<0.05] and no difference between these two groups. SD value of aortic root and pulmonary artery of group Flash 3rd and Sequence 3rd were significantly lower than that of group Flash 2nd( H= -40.27-33.38, P<0.05). SNR of aortic root and pulmonary artery of group Flash 3rd was significantly higher than that of group Flash 2nd and Sequence 3rd ( H=-0.90-51.42, P<0.05). Diagnosis accuracy of intracardiac malformation for group Flash 2nd was significantly lower than that of Flash 3rd and Sequence 3rd (77.7%, 90.9%, 88.9%, K=9.36, P<0.05), and there was no significant difference between the latter two groups. There was no difference in diagnosis accuracy of extracardiac malformation among 3 groups (88.6%, 94.8%, 92.2%, K=3.11, P=0.21). Conclusions:The prospective ECG-gated high-pitch scanning in third generation DSCT can take into account radiation dose and image quality, which has important clinical value in the diagnosis of CHD.