Study on the value of dual energy index and atomic number combined with slope of energy spectrum curve in identifying components of urinary tract stones based on dual energy CT
10.3969/j.issn.1672-8270.2024.03.011
- VernacularTitle:基于双能量CT中双能量指数、原子序数结合能谱曲线斜率对于鉴别泌尿系结石成分的价值研究
- Author:
Ning SUN
1
;
Qilei QIU
;
Yuxue WU
;
Hongli QIAN
;
Yahong BAO
Author Information
1. 南通市海门区人民医院CT室 南通 226100
- Keywords:
Dual energy compute tomography(DECT);
Dual energy index(DEI);
Atomic number;
Slope of energy spectrum curve;
Urinary tract stones
- From:
China Medical Equipment
2024;21(3):53-57
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To assess a value of dual energy index(DEI)and effective atomic number(Zeff)of dual-energy computed tomography(DECT)combined with slope of energy spectrum curve in identifying the component of urinary tract stones.Methods:The clinical and DECT imaging data of 111 patients with urinary tract stones who admitted to Nantong Haimen People's Hospital from October 2019 to October 2022 were selected,and the diameters of the urinary tract stones of all patients were<5 mm.The components of outside body stones of all patients were analyzed by infrared spectrum.The accuracy,sensitivity and specificity of the DEI,Zeff,slope of energy spectrum curve and the combination of the them were respectively calculated in identifying the components of urinary tract stones.Results:In 111 patients,the patients with calcium oxalate stones were 75(67.57%),and ones with hydroxyapatite were 15 cases(13.51%),and ones with uric acid stones were 21 cases(18.92%),and there were no other type of stone included cysteine.The differences of CT values(F=487.691,P<0.001),DEI values(F=395.553,P<0.001),Zeff values(F=818.689,P<0.001)and the slopes of energy spectrum curves(H=19.615,P<0.001)were statistically significant among the three types of stones.The overall accuracies of DEI,Zeff and slope of energy spectrum curve were 87.39%(97/111),84.68%%(94/111)and 81.98%(91/111)in identifying the components of urinary tract stone.The range of diagnostic accuracies of DEI,Zeff and the slope of energy spectrum curve was between 80%and 100%for calcium oxalate stones,hydroxyapatite and uric acid stones.The overall accuracy of DEI,Zeff combined with slope of energy spectrum curve was 94.59%(105/111)in identifying the components of urinary tract stones.The diagnostic accuracies of DEI,Zeff combined with slope of energy spectrum curve were respectively 94.59%,94.59 and 100%for calcium oxalate stones,hydroxyapatite and uric acid stones.Conclusion:Compared with the single indicator of DECT,the DEI and Zeff combined with the slope of the energy spectrum curve showed better diagnostic accuracy in identifying the components urinary track stones.