Analysis of causes and solutions for vacuum suction weakness of dental units
10.3969/j.issn.1672-8270.2024.02.037
- VernacularTitle:口腔综合治疗台负压无力原因分析及解决对策
- Author:
Xinya LI
1
;
Yichen WANG
;
Yansong LIU
;
Anjia ZHENG
;
Shubin WU
;
Baolin FAN
;
Jianxia WANG
Author Information
1. 北京大学口腔医学院(口腔医院)国家口腔医学中心 国家口腔疾病临床医学研究中心 口腔生物材料和数字诊疗装备国家工程研究中心医学装备处 北京 100081
- Keywords:
Vacuum suction;
Weakness;
Dental unit;
Cause analysis;
Countermeasures
- From:
China Medical Equipment
2024;21(2):193-195,199
- CountryChina
- Language:Chinese
-
Abstract:
An efficient vacuum suction system is a necessary prerequisite for the smooth operation of the oral diagnosis and treatment.During the use of the dental units,there is often a situation of vacuum suction weakness,resulting in the inability to discharge the mixture of blood,saliva,dental tissue and other mixtures in time,which affects the doctor's treatment field and increases the risk of aspiration pneumonia and cross-infection in patients.The working principle,pipeline system,filters and other aspects of the vacuum suction system that may affect the suction efficiency was analyzed.The causes and solutions of vacuum suction weakness were discussed,and operation suggestions were proposed to ensure the safe and effective use of equipment and ensure the safety of diagnosis and treatment.