Research on the value of time control mode in the operation management of blood separator
10.3969/j.issn.1672-8270.2024.02.028
- VernacularTitle:时间控制模式在血液分离机运行管理中的价值研究
- Author:
Yali WU
1
;
Jie CHENG
;
Lixia CHEN
;
Yi LIAO
Author Information
1. 上海市徐汇区中心医院医学装备部 上海 200031
- Keywords:
Time control;
Blood separation;
Data mining;
Blood donation experience;
Equipment quality
- From:
China Medical Equipment
2024;21(2):143-148
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To establish a time control mode for component blood collection and to discuss its application value in the operation management of blood separator.Methods:The causes for delay in blood separation time in blood component collection were analyzed,the correlation between influencing factors and time control level was analyzed by using data mining algorithm,and time control management countermeasures were developed by running time prediction.A total of 600 blood donors and 17 blood separators in clinical use in Shanghai Xuhui District Central Hospital from 2019 to 2022 were selected and divided into a control group(300 blood donors,14 blood separators)and an observation group(300 blood donors,15 blood separators,including 12 of control groups and 3 newly added)according to different management modes.The control group adopted the quality control mode,the observation group adopted the time control mode.The number of problems,the proportion of delay,the number of satisfied people and the assessment results of relevant medical staff in the blood component collection process were compared between the two groups.Results:The number of blood donors with problems related to equipment,medical consumables,medical staff and blood donors during the blood component collection process in the observation group was 11(accounting for 3.7%),17(accounting for 5.7%),7(accounting for 2.3%)and 18(accounting for 6.0%),respectively,which were lower than those in the control group,the difference was statistically significant(x2=7.192,6.822,8.153,11.405,P<0.05).The number of blood donors whose blood separators were in normal operation,apparent delay and significant delay in the observation group were 236(accounting for 78.7%),46(accounting for 15.3%)and 18(accounting for 6.0%),respectively,and the proportions of normal operation were higher than those in the control group,the proportions of apparent delay and significant delay were lower than those in the control group,the difference was statistically significant(x2=17.821,7.131,8.549,P<0.05).The proportion of blood donors satisfied was 284(accounting for 94.7%)in the observation group,which was higher than that in the control group,the difference was statistically significant(x2=30.171,P<0.05).The assessment score of medical staff involved in blood donors in the observation group was(91.14±4.86),which was higher than that in the control group,the difference was statistically significant(t=4.587,P<0.05).Conclusion:Time control mode can control the probability of the problems in the operation of blood separator,shorten the delay time of blood separator operation,improve blood donation experience of blood donors and blood collection service quality,and improve the professional level of medical staff.