Research on MRE quality control in diagnosing intestinal diseases
10.3969/j.issn.1672-8270.2024.02.002
- VernacularTitle:磁共振小肠造影技术在小肠疾病诊断中的图像质量控制研究
- Author:
Chujie CHEN
1
;
Zhen CHEN
;
Chaoshang LIN
;
Chengkun HONG
;
Peiyun YE
;
Jiamin CHEN
;
Yonggang LIANG
;
Liyuan FU
Author Information
1. 福建中医药大学福总教学医院(第九〇〇医院)放射诊断科 福州 350025
- Keywords:
Magnetic resonance enterography(MRE);
Examination skill;
Image quality;
Quality control
- From:
China Medical Equipment
2024;21(2):7-11
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To investigate the quality control of magnetic resonance enterography(MRE)in the diagnosis of intestine diseases,and analyze the factors that affected the imaging quality of MRE,and enhance the imaging quality of MRE through adopted the measures of quality control.Methods:The documents of MRE examinations of 167 patients with intestinal disease who admitted to the 900th Hospital of People's Liberation Army Joint Service Support Force from May 2018 to March 2023 were retrospectively analyzed.The image qualities of all patients were evaluated after they completed clinical and image examinations.The reasons that image quality could not meet the requirement of diagnosis were analyzed.And then,the measures of quality control were proposed.Results:In 167 patients with intestinal disease,the MRE images of 153 patients(91.62%)could meet the requirement of diagnosis.In 14 patients(8.38%)whose MRE images could not meet the requirement of diagnosis,the reason of 3 cases(1.80%)was poor respiratory coordination,and that of 2 cases(1.20%)was there were more severe magnetic sensitive artifacts in images,and that of 1 case(0.60%)was severe intestinal peristalsis leaded to blurred images,and that of 2 cases(1.20%)was the flow void effect from intestinal peristalsis inside of intestinal cavity could not meet the requirement of diagnosis,and that of 4 cases(2.40%)was the intestinal tube without incomplete dilation caused by poor oral filling contrast agent,and that of 2 cases(1.20%)was many residues in intestine due to poor preparation for intestine.Aimed at the factors that MRE images could not meet requirement of diagnosis,we proposed the following quality control measures:①the biphasic contrast agents with favorable safety,without severe adverse reactions,which can fully dilate intestinal cavity,should be selected.②we should do well for the dilation of intestinal tube,and inhibit the intestinal peristalsis and conduct respiratory training.③we should conduct scan with wide field at coronal site,so as to display panorama image of intestine.④The scans of conventionally anatomical sequence and functional imaging sequence on axis position were performed on lesions.Conclusion:MRE technique should choose appropriate contrast agent in the quality control of the diagnosis of intestine diseases,and do well the preparation for patients before examination.Using intraluminal contrast agents,conducting intestinal dilation and optimal imaging technique are essential for obtaining intestinal MRE images with high quality.