Atrial fibrillation detection using millimeter-wave radar
10.3969/j.issn.1005-202X.2024.01.012
- VernacularTitle:基于毫米波雷达的心房颤动检测
- Author:
Hengji ZHOU
1
;
Yihan YANG
;
Yuanhui HU
;
Yuguang CHU
;
Xintian SHOU
;
Yaping YOU
;
Wenjing XUE
;
Shaowei FAN
;
Yong WANG
;
Huiliang SHEN
Author Information
1. 浙江大学工程师学院,浙江杭州 315000
- Keywords:
atrial fibrillation;
millimeter-wave radar;
signal processing;
non-contact;
real-time
- From:
Chinese Journal of Medical Physics
2024;41(1):81-87
- CountryChina
- Language:Chinese
-
Abstract:
A novel technology is proposed for non-contact and real-time detection of atrial fibrillation using millimeter-wave radar.A 60 GHz PCR millimeter wave radar is used to continuously detect the chest echo signal of the subject.After signal acquisition,I-Q signal is generated through I-Q demodulation,and the signal phase information is extracted using effective points phase trend evaluation for obtaining the signals from oscillations in the chest wall,from which the respiratory signals and cardiac signals are extracted through digital filtering for the analysis of cardiac movement.Whether the atrial fibrillation occurs or not is determined by the characteristics of atrial fibrillation wave in the time domain.The effective points phase trend evaluation for extracting more accurate signal phase information and the time-domain method for real-time atrial fibrillation detection are the innovations of the study.The experimental results show that the proposed method achieves a detection accuracy of 99.2%in clinic.