Association Between V1 Lead r'Wave Amplitude and Impedance Changes With Left Bundle Branch Pacing Electrode Implantation Depth
10.3969/j.issn.1000-3614.2024.03.010
- VernacularTitle:V1导联r'波振幅结合阻抗变化评价左束支区域起搏电极植入深度的研究
- Author:
Qianghui HUANG
1
;
Yueping JIANG
;
Biming ZHAN
;
Qianwei HUANG
;
Jinzhu HU
Author Information
1. 南昌大学第二附属医院 心内科,南昌 330000
- Keywords:
left bundle branch area pacing;
r'wave amplitude;
electrode perforation;
impedance
- From:
Chinese Circulation Journal
2024;39(3):273-278
- CountryChina
- Language:Chinese
-
Abstract:
Objectives:To explore the association between the r'wave amplitude in lead V1 and impedance changes with left bundle branch pacing electrode implantation depth. Methods:A total of 78 patients with normal heart structure and underwent left bundle branch area pacing(LBBAP)in the Second Affiliated Hospital of Nanchang University from January 1,2019 to December 31,2021 were included in this retrospective analysis.Baseline data,intraoperative and imaging data,and 3,6,9 and 12 months of follow-up results were collected.Correlation and regression analysis were performed to define the feasibility using the r'wave in lead V1 during pacing and impedance changes to estimate the electrode depth. Results:r'waves at the end of the QRS complex in lead V1 during pacing were found in 70 cases(89.7%),and 8 cases(10.3%)showed rS,RS type QRS waves,or no r'wave at the end.Correlation analysis showed that r'wave amplitude was positively correlated with electrode depth(r=0.424,P<0.01),negatively correlated with impedance(r=-0.256,P=0.03).There was no significant statistical correlation between electrode implantation depth and impedance(r=-0.132,P=0.27).Regression analysis found that electrode depth was an important factor affecting r'wave amplitude(regression coefficient=0.056,P=0.000).Combined with the established regression model and impedance,it was found that the amplitude of the r'wave in lead V1 is at the range of 0.24-0.69 mV,and the impedance ranges from 648.30 to 828.90 Ω,and the electrode implantation depth is 6-11 mm,which is most suitable.The risk of perforation is low,and the left bundle branch can be successfully captured with a high probability.The pacing parameters are satisfactory,and the pacing QRS wave duration is narrow.During the intraoperative,postoperative 48 hours,and 12-month follow-up period,the patient did not experience complications such as electrode perforation,thromboembolism,cardiac tamponade,infection,or wire dislocation. Conclusions:Left bundle branch region pacing is a safe and feasible pacing method.During LBBAP,the amplitude of the r'wave in lead V1 at the range of 0.24-0.69 mV,and the impedance ranges from 648.30 to 828.90 Ω can be used to guide the pacing in the left bundle branch region and reduce the risk of electrode perforation.