Efficacy of CARTO Three-dimensional Mapping and Zero X-ray Radiofrequency Catheter Ablation of Ventricular Premature Contraction Originating From Aortic Sinus
10.3969/j.issn.1000-3614.2024.03.009
- VernacularTitle:CARTO三维标测零射线下射频导管消融术治疗主动脉窦起源室性早搏的临床研究
- Author:
Jingyu TIAN
1
;
Zhengyu ZHU
;
Ruizheng WANG
;
Yunli TIAN
;
Jing HUANG
;
Yan WANG
Author Information
1. 华中科技大学同济医学院,武汉 430030
- Keywords:
zero X-ray;
radiofrequency ablation;
aortic sinus;
ventricular premature contract
- From:
Chinese Circulation Journal
2024;39(3):267-272
- CountryChina
- Language:Chinese
-
Abstract:
Objectives:To finely divide the aortic sinus into sections and accurately localize the coronary ostium through CARTO three-dimensional mapping,and to assess the clinical effects of treating aortic sinus cusps premature ventricular contraction(ASC-PVC)and the ablation risk in the corresponding area with zero X-ray radiofrequency ablation. Methods:A total of 66 patients who underwent radiofrequency ablation for ASC-PVC from January 2020 to January 2023 were included in this analysis,patients were divided into experimental group(n=34)and conventional group(n=32).In the conventional group,the CARTO 3 system was used to create an aortic sinus model through the conventional method.The earliest stimulating target was identified by using electrical stimulation mapping(ESM).Radiofrequency ablation treatment was performed after the distance between the target and the coronary ostium was precisely measured by coronary angiography through the hollow tube of the ablation catheter or coronary angiography tube.In the experimental group,the CARTO 3 system was used to build a model of the aortic sinus and the coronary ostium and aortic sinus were divided into sections.The earliest stimulating target was identified by ESM.After localizing coronary ostium through the impedance changing pattern on the ablation catheter tips,catheter ablation was performed with zero X-ray.The data regarding the impedance of the ablation catheter in the aortic sinus were collected.The total operative time,the operative time in the aortic sinus,contrast dosage,X-ray exposure time,immediate and short-term success rates of the operation and complication rates were compared between the two groups.Besides,the distribution of successfully ablated targets and their relationship with the risk of ablation were analyzed in both groups. Results:There was no significant statistical difference in the immediate and short-term success rates between the two groups(93.8%vs.94.1%;90.6%vs.88.2%,both P>0.05).The experimental group did not receive contrast agents during the operation,and the total operation time and intra-aortic sinus operation time in the experimental group were significantly shorter than those in the conventional group([58.76±4.94]min vs.[66.91±5.94]min,P<0.001;[43.12±4.49]min vs.[50.31±5.18]min,P<0.001).During the process of moving the ablation catheter from the intra-aortic sinus to the coronary artery opening and into the coronary artery,the impedance suddenly increased,which was significantly different from the impedance in other parts of the intra-aortic sinus(all P<0.001). Conclusions:Radiofrequency ablation of ASC-PVC with zero X-ray can simplify the procedures and shorten the operative time.The steep increase in impedance at the tip of the ablation catheter can be used as a basis for localizing the coronary ostium.Dividing the aortic sinus into sections allows a detailed assessment of the risk for ablation treatment at the targets.