Common failure analysis and maintenance management strategy for transcranial magnetic stimulator
10.19745/j.1003-8868.2023194
- VernacularTitle:经颅磁刺激仪常见故障分析与维修管理策略
- Author:
Xing-Guang ZHU
1
;
Li-Juan BAI
;
Meng-Fan JIA
;
Xiao-An WANG
Author Information
1. 北京回龙观医院,北京 100096
- Keywords:
transcranial magnetic stimulator;
troubleshooting;
daily maintenance;
utilization management
- From:
Chinese Medical Equipment Journal
2023;44(9):110-113
- CountryChina
- Language:Chinese
-
Abstract:
The working principle and structural composition of the transcranial magnetic stimulator were introduced.The failures of the transcranial magnetic stimulator in some hospital were summarized from 2018 to 2022.The causes for common failures of the transcranial magnetic stimulator were analyzed with the fishbone diagram,and some suggestions were put forward on its daily utilization management.The maintenance thoughts for two common failures were described in detail including flow control alarm and no magnetic field output.References were provided for daily management and maintenance of the transcranial magnetic stimulator.[Chinese Medical Equipment Journal,2023,44(9):110-113]