Construction and effect evaluation of management process of Da Vinci robotic surgical instrument based on bar code
10.3760/cma.j.cn211501-20220826-02711
- VernacularTitle:基于条形码的达芬奇手术器械管理流程构建及效果评价
- Author:
Chen QIAN
1
;
Liuliu ZHANG
;
Tianfu LI
;
Yun NING
;
Zhu YU
;
Ping ZHU
Author Information
1. 江苏省肿瘤医院 江苏省肿瘤防治研究所 南京医科大学附属肿瘤医院手术室,南京 210009
- Keywords:
Da Vinci robotic surgical instrument;
Bar code management;
Reusable medical devices
- From:
Chinese Journal of Practical Nursing
2023;39(30):2341-2345
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To establish and effect evaluate of management process of Da Vinci robotic surgical instrument based on bar code, and evaluate its application effect.Methods:This study was a randomized controlled trial using convenient sampling method. A total of 172 Da Vinci robotic surgeries from Jiangsu Cancer Hospital from September to November 2020 were selected as the control group, and 118 Da Vinci robotic surgeries from September to November 2021 were selected as the observation group. The control group adopted the traditional Da Vinci surgical instrument management mode, while the observation group adopted the management mode of Da Vinci surgical instrument based on bar code. The incidence of management problems related to Da Vinci surgical instruments, the number of reported adverse events of instruments rate and the impact of staff work effect were compared between two groups.Results:The incidence rates of non-traceable information, counting error, usage charge and bar code irregularity were 9.3% (11/118), 0, 0.8% (1/118) and 2.5% (3/118), respectively, which were lower than 26.7% (46/172), 17.4% (30/172), 12.2% (21/172) and 16.9% (29/172) of the control group. The differences were statistically significant ( χ2 values were 12.89-24.68, all P<0.01). The reporting rate of device adverse events in the control group was 0, and the observation group was 4.2% (5/118), and the difference between the two groups was statistically significant ( χ2=7.42, P<0.05). The number of sterilization package turnover sets in the observation group was 2(2, 3), which was lower than 2(2, 4) of the control group, and the difference was statistically significant ( Z=-2.50, P<0.05). The average amount of pre-purchased surgical instruments per surgical instrument in the observation group was 7 858.0 (6 712.1, 14 119.9), which was lower than that of the control group of 7 858.0 (7 858.0, 2 0385.0), and the difference was statistically significant ( Z=-2.97, P<0.05). Conclusions:The management process of Da Vinci Xi robotic surgical instrument based on bar code can reduce the incidence of related management problems, increase the reporting rate of adverse events, improve the work efficiency of the management of mechanical arm to optimize hospital of reusable medical equipment management level, thus further ensure medical safety and improve medical quality.