Establishment and verification of microscopic review rules in Dirui GMD-S600 gynecological secretion analysis system
10.3969/j.issn.1673-4130.2024.02.021
- VernacularTitle:迪瑞GMD-S600妇科分泌物分析系统显微镜复检规则制订及验证
- Author:
Wanchao LIU
1
;
Longmei CHEN
;
Liangxiao ZHU
;
Qingqing LIU
;
Yiqiao DU
Author Information
1. 上海市宝山区中西医结合医院检验科,上海 201900
- Keywords:
vaginal secretions;
review rule;
review rate
- From:
International Journal of Laboratory Medicine
2024;45(2):229-233
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish and verify the microscopic review rules of Dirui GMD-S600 gynecologi-cal secretion analysis system.Methods A total of 2 619 samples of vaginal secretions were collected and test-ed for tangible components and dry chemistry items using the Dirui GMD-S600 gynecological secretions analy-sis system.All samples were subjected to manual microscope examination using a double blind method,and the results of instrument testing,manual image review of instrument testing,and manual microscope examina-tion were recorded separately.Using the results of manual microscope examination as the standard,the false negative rates and the false positive rates of instrument testing results and tangible component results after manual image review of instrument testing were analyzed.At the same time,the false negative rates and the false positive rates of instrument detection dry chemistry items and their corresponding tangible components were analyzed,and the review rules were derived based on this.Subsequently,344 samples were selected to verify the effectiveness of the review rules.Results The false positive rates of white blood cells,fungi,trichomonas and clue cells detected by the instrument were 28.77%,14.71%,20.09%and 2.30%,and the false negative rates were 0.28%,15.06%,4.35%and 10.34%.After manual image review and correction,the false positive rates of white blood cells,fungi,trichomonas and clue cells were 28.77%,7.06%,8.67%,0.66%,and the false negative rates were 0.28%,3.35%,4.35%and 10.34%.The false positive rates of white blood cell esterase(LE)for white blood cells,N-acetylhexylglucosidase(NAG)for fungi,NAG for trichomonas,hydrogen peroxide(H2O2)for bacteria,lactic acid(LA)for bacteria,sialidase(SNA)for clue cells,and proline aminopeptidase(PIP)for clue cells were 13.83%,22.47%,24.10%,95.52%,2.52%,12.92%and 5.54%,and the false negative rates were 28.17%,48.12%,0.00%,23.55%,78.82%,0.00%and 74.14%.The false positive rate of 260 validation samples was 1.68%,the false negative rate was 4.22%,and the review rate was 45.38%.Conclusion The review rules of the Dirui GMD-S600 gynecological secretion analysis system are as follows(1)All specimens need to be reviewed with images,and the following rules are based on this rule.(2)When the result of white blood cell shows positive,then the microscopic exam-ination is required.(3)When the result of fungi shows positivity,then the microscopic examination is re-quired.(4)When the result of trichomonas shows positive,then the microscopic examination is required.(5)When the result of SNA does not match the positive result of clue cells,then the microscopic examination is required.