Consistency evaluation between liquid chromatography tandem mass spectrometry and Electrochemiluminescence immunoassay in detecting 25-hydroxyvitamin D
10.3969/j.issn.1673-4130.2023.24.021
- VernacularTitle:液相色谱串联质谱法和电化学发光免疫分析法检测25-羟维生素D的一致性评价
- Author:
Ke ZHAO
1
;
Weidan REN
;
Yuxia LIU
;
Wenbing YUE
;
Xueying HAN
;
Lixia DING
Author Information
1. 郑州金域临床检验中心有限公司,河南郑州 450000
- Keywords:
liquid chromatography-tandem mass spectrometry;
electrochemiluminescence immunoassay;
25-hydroxyvitamin D
- From:
International Journal of Laboratory Medicine
2023;44(24):3050-3053,3058
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the correlation and consistency of liquid chromatography-tandem mass spectrometry(LC-MS/MS)and electrochemiluminescence immunoassay(ECLIA)in the determination of se-rum 25-hydroxyvitamin D[25(OH)D],in order to guide the clinical selection of appropriate detection meth-ods.Methods A total of clinical serum samples were collected from the laboratory,and 25(OH)D levels were detected by LC-MS/MS and ECLIA,respectively.Passing-Boblok regression was used to analyze the correla-tion between the two methods,and Bland-Altman and Mountain plot were used to evaluate the agreement be-tween the two methods.Serum 25(OH)D<20.0 ng/mL was defined as vitamin D deficiency,and serum 25(OH)D as 20.0-<30.0 ng/mL was defined as vitamin D insufficiency.Kappa analysis was used to determine the coincidence rate of the two methods in the diagnosis of vitamin D nutritional status.Results The 25(OH)D levels detected by LC-MS/MS and ECLIA were(26.67±4.79)ng/mL and(39.33±4.09)ng/mL,respec-tively.The regression equation of the two methods was YECLIA=-4.558 1+1.719 8XLC-MS/MS,the slope was 1.719 8(95%CI 1.586 3-1.828 4),excluded 1,and the intercept was 4.558 1(95%CI-7.692 2--2.122 1),excluded 0,prompt system difference or ratio differences of the two methods.There were system-atic or proportional differences between the two methods.The Bland-Altman figure showed two methods aver-ages was 12.7,and the difference of out points(ratio)was 3.19%.The peak value of the mountain plot was-9.17 ng/mL,with more deviations from 0,indicating poor agreement between the results measured by the two methods.The Kappa coefficient of the two methods for judging vitamin D deficiency was 0.875,and the diagnostic coincidence rate was 94.68%.The Kappa coefficient of the two methods for judging vitamin D in-sufficiency was 0.538,and the diagnostic coincidence rate was 75.53%.Conclusion The agreement between ECLIA and LC-MS/MS is poor,but the agreement between ECLIA and LC-MS/MS in the diagnosis of vita-min D deficiency and insufficiency based on nutritional status is high.