Evaluation of HbA1c Levels Via the Latex Immunoturbidimetric Method by Using Chemistry Autoanalyzer.
- Author:
Yongjun JO
1
;
So young LEE
;
Hae il PARK
;
YeongSic KIM
;
Jehoon LEE
;
Yonggoo KIM
;
Kyungja HAN
Author Information
- Publication Type:Original Article
- Keywords: HbA1c; Immunoturbidimetric method; HPLC
- MeSH: Chromatography, High Pressure Liquid; Chromatography, Liquid; Diabetes Mellitus; Hemoglobins; Hemolysis; Latex; Organothiophosphorus Compounds; Water
- From:Laboratory Medicine Online 2012;2(1):10-14
- CountryRepublic of Korea
- Language:Korean
- Abstract: BACKGROUND: Measurement of HbA1c levels is widely used to diagnose diabetes mellitus and to evaluate and monitor plasma-glucose concentrations over 6-8 weeks. In this study, we evaluated the diagnostic performance of the newly developed latex immunoturbidimetric method by using Autolab HbA1c. METHODS: We analyzed and compared the diagnostic performance of Autolab HbA1c with that of Toshiba 200FR between April 2009 and July 2009. According to guidelines (EP5-A2, EP6-P, EP9-A2) of the clinical and laboratory standards institute (CLSI), we compared linearity, precision and correlation of Autolab HbA1c with those of G7 (Tosoh Corp., Kyoto, Japan) by using high-performance liquid chromatography (HPLC) method. RESULTS: Data obtained using Autolab HbA1c showed good linearity in mixtures of samples with low (3.1%) and high (15.1%) levels of HbA1c (r2 = 0.9997). In the analysis of within-run precision of the samples with HbA1c levels of 5.1% and 12.1%, the SDs were 0.04 and 0.06 and covariances of these samples were 0.8% and 0.5%, respectively. In the Deming regression model, the regression equation was as follows: Autolab HbA1c = 1.0859xTosoh HPLC-0.6957. CONCLUSIONS: In this study, Autolab HbA1c method showed better performance characteristics than Tosoh G7 did. In reference review, there was no interference of variant hemoglobin. The data acquisition time of Autolab HbA1c was lower than that of Tosoh G7. The advantages of Autolab HbA1c are that it can be used as an autoanlyzer in routine chemical analysis, it does not require pre-analytical treatment, and the samples are automatically treated with distilled water for hemolysis.