Repeatability of wavefront aberration measured by adaptive optics visual simulator and agreement of OPD-Scan Ⅲ
10.3980/j.issn.1672-5123.2024.5.28
- VernacularTitle:自适应光学视觉模拟器测量波前像差的重复性及与OPD-Scan Ⅲ的一致性研究
- Author:
Yan WANG
1
;
Xuan LIAO
1
;
Changjun LAN
1
;
Biao LI
1
;
Zhuang MIAO
1
;
Qingqing TAN
1
;
Suyun QIN
1
;
Huan HUANG
1
Author Information
1. Department of Ophthalmology, Affiliated Hospital of North Sichuan Medical College, Nanchong 637000, Sichuan Province, China
- Publication Type:Journal Article
- Keywords:
adaptive optics visual simulator;
OPD-Scan Ⅲ;
wavefront aberration;
repeatability;
agreement
- From:
International Eye Science
2024;24(5):810-815
- CountryChina
- Language:Chinese
-
Abstract:
AIM: To assess the repeatability and agreement of higher-order aberration obtained by adaptive optics visual simulator(VAO)compared with OPD-Scan Ⅲ.METHODS: A cross-sectional study was conducted from August to September 2023, including a total of 204 patients(204 eyes)with myopia whose right eyes were measured. The examinations were performed by the same skilled examiner using both devices separately. The VAO device was used to measure higher order aberrations of orders 3 to 6 at a pupil diameter of 4.5 mm, while both the VAO and OPD-Scan Ⅲ devices were utilized to measure total higher-order aberration(tHOA), spherical aberration(SA), coma aberration(Coma), and trefoil aberration(Trefoil)of the entire eye at pupil diameters ranging from 3 to 6 mm. Furthermore, the repeatability of whole eye aberration measurements obtained with the VAO device was evaluated and the agreement of the two devices was assessed.RESULTS: The whole-eye higher-order aberrations measured by VAO demonstrated excellent repeatability(0.767≤ICC≤0.941, Sw<0.01 μm, TRT<0.1 μm). There was no statistically significant difference in Coma measured by VAO or OPD-Scan Ⅲ for pupil diameters ranging from 4 to 6 mm(P>0.05), while a statistically significant difference was observed in whole-eye tHOA of other pupil diameters(all P<0.05). The agreement of aberration measurements for each order between VAO and OPD-Scan Ⅲ for 3 mm pupil diameters, SA at 4 and 5 mm pupil diameter and Coma at 4 mm pupil diameter showed a 95% limit of agreement(LoA)<0.1, indicating good agreement; however, poor agreement was found for the remaining aberration measurements at different pupil diameters, with a 95%LoA>0.1, and there were significant differences in higher-order aberrations measured by two devices under a pupil diameter of 3 mm(r=0.218-0.317, P<0.01), 4 mm(r=0.406-0.672, P<0.01), 5 mm(r=0.538-0.839, P<0.01 and r=0.030-0.109, P>0.01)and 6 mm(r=0.369-0.766, P<0.01).CONCLUSION: The VAO demonstrates favorable repeatability when assessing whole-eye higher order aberration under pupil diameters of 3-6 mm. However, there is inadequate agreement and interchangeability in whole-eye higher order aberration at 3-6 mm pupil diameter between VAO and OPD-Scan Ⅲ for clinical purposes.