Practical Optimization and Application of Time-delay Exposure System for Mobile Digital Radiography Equipment.
10.3969/j.issn.1671-7104.2023.06.020
- Author:
Zhihao FU
1
;
Chao DU
1
;
Chuanjun XU
1
;
Yuting TIAN
1
Author Information
1. Department of Radiology, Nanjing Second Hospital, Nanjing University of Chinese Medicine, Nanjing, 210003.
- Publication Type:Journal Article
- Keywords:
X-ray;
mobile digital radiography;
time-delay exposure system
- MeSH:
Radiographic Image Enhancement;
Technology;
Electric Power Supplies
- From:
Chinese Journal of Medical Instrumentation
2023;47(6):695-697
- CountryChina
- Language:Chinese
-
Abstract:
This study introduced a time-delay exposure system independent of the mobile digital radiography equipment. The system consisted of lithium battery, delay control circuit, micro electric motor and related auxiliary facilities. When the starting time was reached through the delay circuit, the motor pushed out the rod to squeeze the exposure button and completed the exposure. The accessories used in this system were easy to purchase and cheap. At the same time, the technology was mature and had good compatibility. The exposure success rate was high and the exposure effect was satisfactory. This time-delay exposure system had good practicability and popularization value.