Three-dimensional-printed hemi-pelvic prosthesis for revision of aseptic loosening or screw fracture of modular hemi-pelvic prosthesis.
10.7507/1002-1892.202306073
- Author:
Xin LIU
1
;
Yi LUO
1
;
Xuanhong HE
1
;
Jie WANG
1
;
Zhuangzhuang LI
1
;
Yuqi ZHANG
1
;
Xin HU
1
;
Minxun LU
1
;
Fan TANG
1
;
Yong ZHOU
1
;
Li MIN
1
;
Chongqi TU
1
Author Information
1. Department of Orthopedics, Orthopedic Research Institute, West China Hospital, Sichuan University, Chengdu Sichuan, 610041,P. R. China.
- Publication Type:Journal Article
- Keywords:
Pelvic tumor;
modular hemi-pelvic prosthesis;
revision;
three-dimesional-printed hemi-pelvic prosthesis
- MeSH:
Male;
Female;
Humans;
Adult;
Arthroplasty, Replacement, Hip/methods*;
Hip Prosthesis;
Treatment Outcome;
Prosthesis Failure;
Fractures, Bone/surgery*;
Retrospective Studies;
Neoplasms
- From:
Chinese Journal of Reparative and Reconstructive Surgery
2023;37(10):1183-1189
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:To investigate the effectiveness of three-dimensional (3D)-printed hemi-pelvic prosthesis for revision of aseptic loosening or screw fracture of modular hemi-pelvic prosthesis.
METHODS:Between February 2017 and January 2020, 11 patients with aseptic loosening or screw fracture of modular hemi-pelvic prosthesis were revised using 3D-printed hemi-pelvic prostheses. There were 7 males and 4 females with an average age of 44 years (range, 25-60 years). In the first operation, all patients underwent total tumor resection, modular hemi-pelvic prosthesis reconstruction, and autologous femoral head transplantation. According to the Enneking pelvic partition system, 8 cases were resected in zones Ⅰ+Ⅱ and 3 cases in zones Ⅰ+Ⅱ+Ⅲ. The interval from the initial operation to this revision ranged from 14.3-66.2 months, with an average of 35.8 months. The operation time, the amount of intraoperative bleeding, and the occurrence of complications were recorded. At 6 months after the first operation, before revision, and at last follow-up, the American Musculoskeletal Tumor Society (MSTS) score and Harris score were used to evaluate the recovery of lower limb function. The pain-free walking distance of patients without brace assistance was recorded at last follow-up. X-ray films were taken at 1 month after the first operation, before revision, and at 1 month after revision, the acetabulum position was assessed by the differences in weight arm and cup height between bilateral hip joints. At last follow-up, the digital X-ray tomography was taken to evaluate the prosthesis-bone integration and the occurrence of aseptic loosening.
RESULTS:The operation time was 182.6-238.0 minutes (mean, 197.4 minutes). The amount of intraoperative bleeding was 400-860 mL (mean, 550.0 mL). All incisions healed by first intention with no infection, hip dislocation, nerve damage, or vascular-related adverse events. The MSTS score and Harris score at last follow-up were significantly higher than those at 6 months after the first operation and before revision ( P<0.05), while the score before revision was significantly lower than that at 6 months after the first operation ( P<0.05). At last follow-up, the patients were able to walk more than 1 000 meters painlessly without brace assistance. Imaging review showed that the difference of cup height at 1 month after revision was significantly lower than that at 1 month after the first operation and before revision, and at 1 month after the first operation than before revision operation, and the differences were significant ( P<0.05). There was no significant difference in the difference of weight arm among three time points ( P>0.05). All prostheses were well integrated, and no aseptic loosening of the prosthesis or screw fracture occurred.
CONCLUSION:Revision with 3D-printed hemi-pelvic prostheses benefited in reconstructing stable pelvic ring and natural bodyweight transmission for patients encountering the aseptic loosening or screw fracture of modular hemi-pelvic prosthesis. Early postoperative rehabilitation training can maximize the recovery of patient limb function, reduce pain during walking, and reduce the incidence of complications.