An in vitro study of temperature variation of super-pulsed thulium fiber laser in ureter
10.3969/j.issn.1009-8291.2023.11.016
- VernacularTitle:超脉冲铥光纤激光在输尿管中温度变化的体外实验
- Author:
Tianfu DING
1
;
Bo XIAO
1
;
Zhongyue HUANG
1
;
Xue ZENG
1
;
Lei LIANG
1
;
Chaoyue JI
1
;
Jianxing LI
1
Author Information
1. Department of Urology, Beijing Tsinghua Changgung Hospital, Tsinghua University, Beijing 102218, China
- Publication Type:Journal Article
- Keywords:
super-pulsed thulium laser;
ureteral lithotripsy;
thermal effect;
safety threshold
- From:
Journal of Modern Urology
2023;28(11):993-997
- CountryChina
- Language:Chinese
-
Abstract:
【Objective】 The thermal effects of super-pulsed thulium fiber laser (TFL) at different powers,lithotripsy modes and irrigation rates were studied using a 3D kidney model to simulate ureteral lithotripsy in vivo. 【Methods】 A thermal effect model was established in vitro. Under the same conditions of laboratory temperature and equipment,the temperature around the optical fiber was measured and compared when different optical fiber diameters,powers,lithotripsy modes and irrigation rates were used to simulate lithotripsy by TFL. 【Results】 There was significant difference in the temperature around the optical fiber caused by two fibers with different diameters under the same conditions (P<0.05). Under the same conditions,different lithotripsic modes produced different temperatures,and the temperature of "high energy and low frequency" was lower than that of "low energy and high frequency" (P<0.05). When the power was 10 W and the minimum irrigation rate was 10 mL/min,the plateau temperature did not reach the safety threshold (43 ℃). When the power was 20 W and the minimum irrigation rate was 10 mL/min,the platform temperature exceeded the safety threshold. When the irrigation rate was 20 mL/min,the platform temperature did not reach the safety threshold. 【Conclusion】 In the study of ureteral lithotripsy in vitro,the power,mode,irrigation rate and optical fiber diameter are factors affecting the thermal effects of TFL. No matter what kind of lithotripsy mode and fiber diameter,the temperature around the fiber is safe when the lithotripsy power is ≤10 W and the irrigation rate is ≥10 mL/min;when the lithotripsy power is ≤20 W and the irrigation rate is ≥20 mL/min,the temperature around the fiber is safe.