Comparison of image quality, diagnostic accuracy, and LV parameters in filtered backprojection versus iterative SPECT Reconstruction in Myocardial Perfusion Scintigraphy
- Author:
Deverly D. Tumapon
1
;
Jerry M. Obaldo
1
Author Information
1. Division of Nuclear Medicine, Philippine Heart Center
- Publication Type:Journal Article
- Keywords:
Iterative reconstruction;
FBP;
Image quality
- From:
The Philippine Journal of Nuclear Medicine
2022;17(1):24-33
- CountryPhilippines
- Language:English
-
Abstract:
Background:Myocardial perfusion scintigraphy images with iterative reconstruction showed higher qualitative and
quantitative estimates.
Objective:This study was conducted to compare image quality, diagnostic accuracy, and LV parameters with standard FBP versus OSEM as baseline quality assurance for the choice of reconstruction algorithm most applicable in the local setting.
Methods:This was a retrospective cross-sectional study of 55 Thallium-201 (Tl-201) scans and 14 Techentium-99m
sestamibi (Tc-99m) images that were reprocessed with filtered backprojection (FBP) and with ordered subset expectation maximization (OSEM). The image quality, diagnostic accuracy, and LV parameters of both SPECT reconstructions were compared.
Results:OSEM reconstruction resulted in a significantly smooth background tracer activity as opposed to the standard FBP in both Tl-201 (p<0.01) and Tc-99m (p=0.01) scans. Distinct LV border was superiorly seen in OSEM. There was no significant difference in the diagnostic accuracy between the two reconstruction parameters despite 2 cases of interpretation discrepancies. End-diastolic volume generated by OSEM was significantly higher (117.60+55.14 vs 118.51+55.11, p<0.01 in Tl-201 and 101.07+75.58 vs 104.29+80.39, p =0.01 in Tc-99m). With Tc-99m scans the end-systolic volume and ejection fraction in OSEM were likewise significantly different from FBP (p=0.04 and p<0.01, respectively).
Conclusion:OSEM reconstruction offers smooth background tracer activity with comparable diagnostic accuracy to FBP. All the LV parameters in OSEM were significantly different from FBP in Tc-99m scans and only the EDV in thallium images.
- Full text:17 (1) article 3.pdf