A Sensitive and Specific Liquid ChromatographyTandem Mass Spectrometry Assay for Simultaneous Quantification of Salivary Melatonin and Cortisol:Development and Comparison With Immunoassays
	    		
		   		
		   			 
		   		
	    	
    	 
    	10.3343/alm.2021.41.1.108
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Sunghwan SHIN
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Hyeonju OH
			        		
			        		;
		        		
		        		
		        		
			        		Hea Ree PARK
			        		
			        		;
		        		
		        		
		        		
			        		Eun Yeon JOO
			        		
			        		;
		        		
		        		
		        		
			        		Soo-Youn LEE
			        		
			        		
		        		
		        		
		        		
    Author Information Author Information
 
			        		
			        		
			        			1. Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea
 
 
- Publication Type:Brief Communication
- From:Annals of Laboratory Medicine
	            		
	            		 2021;41(1):108-113
	            	
            	
- CountryRepublic of Korea
- Language:English
- 
		        	Abstract:
			       	
			       		
				        
				        	 Melatonin and cortisol are clinically important for diagnosing sleep and mood disorders.We developed and validated a liquid chromatography-tandem mass spectrometry (LC-MS/ MS) assay for simultaneous measurement of salivary melatonin and cortisol concentrations according to the Clinical and Laboratory Standards Institute guidelines. Additionally, we compared the LC-MS/MS assay with immunoassays, ELISA (Direct Salivary Melatonin Elisa EK-DSM, Bühlmann Laboratories AG, Schönenbuch, Switzerland) and electrochemiluminescence immunoassay (Cortisol II, Roche, Mannheim, Germany), using 121 saliva samples. The LC-MS/MS assay exhibited good performance in terms of linearity, precision, accuracy, limit of detection, lower limit of quantification, extraction recovery, carry-over, and matrix effect. The LC-MS/MS assay and immunoassays showed strong correlation (Pearson’s r = 0.910 for melatonin, r = 0.955 for cortisol), but demonstrated a significant mean bias of 23.2% (range 54.0–143.7%) for melatonin and 48.9% (range 59.7–184.7%) for cortisol. Our LC-MS/MS assay provided more sensitive and reliable salivary melatonin and cortisol quantification results compared with immunoassays.