A enzymatic method of the determination of calcium in serum
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:尿素酰基裂解酶酶偶联法测定血清总钙
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xingliang JIANG
			        		
			        		;
		        		
		        		
		        		
			        		Zhong TANG
			        		
			        		;
		        		
		        		
		        		
			        		Jun ZHANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Clinical Laboratory Science
	            		
	            		 2001;19(3):145-146
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective A enzymatic method for measurement of calcium in serum by using urea amidolyase was introduction. Methods Double reagent two-point method.TEA buffer(pH 8.0,200 mmol/L)was used as the buffer solution.The first reagent consisted of sodium bicarbonate 26.0 mmol/L,NADPH 0.4 mmol/L,2-oxoglutaric acid 8.0 mmol/L,urea amidolyase 115 U/L and GLDH 43.0 kU/L.The second reagent contained sodium bicarbonate 26.0 mmol/L and ATP 6.2 mmol/L in the buffer solution.Results The Linearity range was 0.5~5.0 mmol/L.The average rate of recovery was 102.7%.The average within-run and between-run CV were 2.13% and 2.69%.Camparison with OCPC(X1),showed that Y1=1.02X1-0.021,r=0.981,and with Arsenazo Ⅲ(X2),Y2=0.99X2+0.03,r=0.978.No interference from as much as 300 μmol/L of bilirubin,4 g/L of hemoglobin,2.5 mmol/L of magnesium,2.0 mmol/L of ammonium ion and 8.0 mmol/L of triglyceride.Conclusion This method was simple,accurate and ripid,suitable for automatic analysis of calcium in serum.